MAGNETOM ESSENZA
Established 1.5T performance. With Tim+Dot.
Low installation and power requirements, ultra short magnet, lightweight design, Zero Helium boil-off technology, Tim+Dot workflow suites and engines, the latest software platform and applications. MAGNETOM ESSENZA – a Tim+Dot system.
Overview Technical Details
Field strength | 1.5 Tesla |
Bore size | 60 cm |
System length | 147 cm |
Minimum room size1 | 30 m2 |
RF technology | |
Maximum number of channels2 | 25, 46 |
Number of independent receiver channels that can be used simultaneously in one single scan and in one single FoV, each generating an independent partial image | 8, 16 |
Gradient strength | V-engine (30 mT/m @ 100 T/m/s) |
Helium consumption | Zero Helium boil-off technology |
Siemens' technologies
- Tim integrated coil technology changed MRI forever. With excellent image quality and fast acquisition speed, the proven technology is now in nearly 10,000 installations. Whole-body coverage means no repositioning for multiple exams and more exams per day.
- Dot is a new way of scanning in MRI – a better way. Dot scanning uses a suite of customizable engines. Allowing the user to personalize exams according to patient needs, build in step-by-step user guidance, and automate MRI exams – either "out of the box" or based on the institution's standards.
More information about Dot and all of the Dot engines
Magnet technology
Superconducting, short bore, patient-friendly design with flared opening. Easy siting due to AS (Active Shielding) and E.I.S. (External Interference Shielding) magnet technology.
Field strength | 1.5 Tesla |
Bore size | 60 cm |
Magnet length | 131 cm |
Helium consumption | Zero Helium boil-off technology |
Gradient System
The combination of our world-class gradients together with Tim+Dot technology enables the most demanding applications.
General Features
- Actively shielded (AS), super-compact whole-body gradient coil system
- Extremely low eddy currents
- Water-cooled coil and amplifier for maximum performance
Excellent spatial and temporal resolution
- Shortest TE and TR for high image contrast and best signal-to-noise ratio
Patient-specific shimming
- Active shim via 3 linear channels
- Includes advanced, nonlinear 2nd order shim
Gradients | V - engine: 30 mT/m |
Slew rate | V - engine: 100 T/m/s |
RF technology
Proven Tim technology on MAGNETOM ESSENZA allows a maximum number of 46 channels (coil elements) that can be connected simultaneously. The system has 16 independent RF channels that can be used simultaneously in one single scan and in one single FOV, each generating an independent partial image.
- Solid-state amplifier, maintenance-free
- Digital quadrature demodulation
- Digital filtering
- Receiver with high dynamic range without adjustments
Siting and installation
MAGNETOM ESSENZA allows a cost effective installation and easy siting.
- Only 2 electronic cabinets
- Save up to 25 %3 on the Installation cost with the light-weight, ultra-short Magnet and no need for a separate equipment room
- Profit from unique features that are good for your financial health, and for the environment too
- Low set-up costs
- Low power requirements
- Low running costs that may decrease your energy bill by up to 50%1 when replacing an existing MRI system
- Maximize uptime with a state-of-the-art zero helium boil-off magnet
System length | 147 cm |
Minimum room size1 | 30 m² |
Related Products, Services & Resources
1Minimum total space requirement for magnet, electronics, and console room.
2Channels (coil elements) that can be connected simultaneously.
3Results may vary. Data on file.